By:
Dave Dlugos
Product Technical Leader
Published on:
September 9th, 2024
Last updated on:
November 10th, 2026
Subscribe now and get the latest blog posts delivered straight to your inbox.
How to Perform an RTD Calibration
By:
Dave Dlugos
Product Technical Leader
Published on:
September 9th, 2024
Last updated on:
November 10th, 2026
An RTD calibration compares a resistance temperature detector’s reading with a calibrated reference thermometer at selected temperatures. The results show how closely the sensor measures temperature and whether it meets the application’s accuracy requirements.
RTDs are used in chemical, refining and other processes that require accurate, repeatable temperature measurements. Understanding the sensor’s resistance characteristics, tolerance class and calibration results helps you evaluate its performance.
Read this article to learn:
- How RTDs measure temperature and how tolerance classes affect accuracy.
- What equipment and steps are involved in an RTD comparison calibration.
- What calibration traceability means and how to interpret a calibration report.
- How to determine an appropriate RTD recalibration interval.
How do RTD temperature instruments work?
An RTD measures temperature through the predictable change in electrical resistance of its sensing element. Platinum is commonly used because its resistance characteristics are stable and well characterized.
A Pt100 element has a nominal resistance of 100 Ω at 0 °C. Actual resistance varies within its specified tolerance; a calibrated RTD does not have to read exactly 100.000 Ω at that temperature.
A transmitter, programmable logic controller (PLC) or distributed control system (DCS) converts the resistance signal into temperature for process monitoring and control. This allows the system to regulate equipment such as control valves, heaters and pumps.
The resistance-to-temperature relationship is approximately linear. The measuring electronics must use the appropriate sensor characteristic to convert resistance into temperature accurately.
Figure 1: RTD Calibration Process.
Figure 2: RTD, Resistance to Temperature Relationship.
What should you specify before calibrating an RTD?
Specify the sensing element, resistance characteristic, tolerance class and operating range before selecting calibration points.
Consider these four factors:
- Select the element material, such as platinum.
- Identify the nominal resistance and compatible measuring electronics.
- Choose the tolerance class needed for the process.
- Confirm the element construction and rated temperature range.
The following tables show the tolerance equations and temperature values from the original chart for Pt100 elements with a temperature coefficient of 0.00385 Ω/Ω/°C.
In the tolerance equations, |t| means the absolute value of temperature in °C. The element’s tolerance range does not necessarily represent the operating limits of the complete probe assembly. Confirm those limits with the manufacturer.
Table 1: Pt100 element tolerances and ranges.
| Class | IEC/DIN designation | Construction | Tolerance (±°C) | Chart range (°C) |
|---|---|---|---|---|
| AA | F0.1 | Thin film | 0.1 + 0.0017 × |t| | 0 to 150 |
| AA | W0.1 | Wire wound | 0.1 + 0.0017 × |t| | −50 to 250 |
| A | F0.15 | Thin film | 0.15 + 0.002 × |t| | −30 to 300 |
| A | W0.15 | Wire wound | 0.15 + 0.002 × |t| | −100 to 450 |
| B | F0.3 | Thin film | 0.3 + 0.005 × |t| | −50 to 600 |
| B | W0.3 | Wire wound | 0.3 + 0.005 × |t| | −196 to 600 |
Table 2: Pt100 tolerance values (±°C).
| Temperature (°C) | F0.1 | W0.1 | F0.15 | W0.15 | F0.3 | W0.3 |
|---|---|---|---|---|---|---|
| −196 | — | — | — | — | — | 1.28 |
| −100 | — | — | — | 0.35 | — | 0.80 |
| −67 | — | — | — | 0.28 | — | 0.64 |
| −50 | — | 0.19 | — | 0.25 | 0.55 | 0.55 |
| −30 | — | 0.15 | 0.21 | 0.21 | 0.45 | 0.45 |
| 0 | 0.1 | 0.1 | 0.15 | 0.15 | 0.3 | 0.3 |
| 100 | 0.27 | 0.27 | 0.35 | 0.35 | 0.8 | 0.8 |
| 150 | 0.355 | 0.355 | 0.45 | 0.45 | 1.05 | 1.05 |
| 250 | — | 0.525 | 0.65 | 0.65 | 1.55 | 1.55 |
| 300 | — | — | 0.75 | 0.75 | 1.8 | 1.8 |
| 350 | — | — | — | 0.85 | 2.05 | 2.05 |
| 450 | — | — | — | 1.05 | 2.55 | 2.55 |
| 500 | — | — | — | — | 2.8 | 2.8 |
| 600 | — | — | — | — | 3.3 | 3.3 |
A dash indicates a temperature outside the range shown for that element designation. Values retain the original chart’s displayed rounding.
Operating beyond the specified temperature range can stress the element, lead to aging effects and affect its subsequent performance.
How is an RTD comparison calibration performed?
Comparison calibration places the RTD and a calibrated reference thermometer in the same controlled temperature source. A high-accuracy platinum resistance thermometer (PRT) commonly serves as the reference. NIST also uses comparison measurements against an ITS-90-calibrated standard PRT for industrial thermometer calibrations.
What equipment do you need?
You need a suitable temperature source, a calibrated reference thermometer and compatible precision measuring equipment.
The setup typically includes:
- A dry-block calibrator or stirred liquid bath covering the required temperature range.
- A reference PRT and its readout.
- A precision resistance meter or suitable RTD readout.
- The sensor specifications and selected calibration points.
Figure 3: Comparison calibration using a reference thermometer and an RTD.
What are the calibration steps?
Measure the RTD and reference at each selected test temperature, then compare their readings.
- Isolate the sensor from the process. Follow the site’s procedures before removing or disconnecting the RTD.
- Position the sensors. Place the RTD and reference thermometer in the bath or dry block at suitable immersion depths.
- Connect the measuring equipment. Use the correct RTD wiring configuration. Disconnect the RTD from the process electronics when testing the sensor separately.
- Establish the test temperature. Set the temperature source to the first calibration point and allow the readings to stabilize.
- Record the readings. Document the reference temperature and the RTD’s resistance or indicated temperature.
- Repeat and compare. Complete the remaining test points. Convert resistance measurements using the applicable resistance-to-temperature relationship and compare the results with the reference.
Calibration points may be specified by the customer or defined by the sensor manufacturer.
What does calibration traceability mean?
Traceability connects a measurement result to a specified reference through a documented, unbroken chain of calibrations, with each calibration contributing to measurement uncertainty. The chain may include national metrology institutes such as NIST.
ISO/IEC 17025 accreditation and metrological traceability are related but distinct. Accreditation addresses laboratory competence within a defined scope; traceability describes the measurement result’s connection to its reference.
Figure 4: Illustration of a calibration traceability chain.

How do you read an RTD calibration report?
An RTD calibration report compares the unit under test (UUT) with the reference at each test point. The sensor may also be called the device under test (DUT) or sensor under test (SUT).
The report documents the calibration results and identifies the relevant measuring equipment and reference sensors.
Table 3: Sample RTD calibration results.
| Test point (°C) | Reference (°C) | UUT 1 (°C) | Reported error (°C) | Lower tolerance limit (°C) | Upper tolerance limit (°C) |
|---|---|---|---|---|---|
| 0 | 0.030 | 0.040 | −0.010 | −0.30 | 0.30 |
| 100 | 100.030 | 100.090 | −0.060 | −0.80 | 0.80 |
| 200 | 200.040 | 200.110 | −0.070 | −1.30 | 1.30 |
This sample calculates its reported “error” as reference temperature minus UUT reading. For example:
100.030 °C − 100.090 °C = −0.060 °C
The listed differences fall within the sample’s tolerance limits. A formal statement of conformity also depends on the applicable decision rule and measurement uncertainty.
Figure 5: Sample RTD differences plotted against tolerance limits.

How often should you calibrate an RTD?
Set the recalibration interval according to the process tolerance, operating conditions, calibration history and applicable quality requirements. Review the interval as performance data becomes available.
Mechanical shock, vibration, elevated temperatures and moisture can affect platinum resistance thermometers. High vibration is a potential source of drift and should be considered when selecting the sensor and establishing its calibration interval.
An RTD failure does not always produce an open circuit; a sensor can also develop a short. Sensor faults and diagnostic coverage should be evaluated as part of the complete measurement system.
Figure 6: Illustrative relationship between drift, process tolerance and recalibration timing.

The six- and 24-month intervals in Figure 5 are examples, not universal requirements for food, pharmaceutical or other industries. Tighter process tolerances and evidence of drift may justify shorter intervals.
For questions about your application, contact an Ashcroft product expert. In the meantime, download our guide to learn more about our Temperature Sensors.
Dave Dlugos, Product Technical Leader
Dave Dlugos has a BSEE degree and 40 years of experience in the measurement industry performing design engineering and product management. He has earned 4 U.S. patents and joined Ashcroft in 2007, currently as the Product Technical Leader. He is a senior member of the International Society of Automation (ISA), past ISA District 1, Vice President, past ISA water and wastewater division board member and the past President of CT Valley ISA Section.